Operation Principle
|
High temperature (680 ˚C) catalytic combustion
|
Measurement Technique
|
Non-dispersive infrared (NDIR) detection
|
Measurement Range
|
100 ppb C – 30,000 ppm C
(multiple calibration ranges or dilution required)
|
Instrument Detection Limit (IDL)
|
50 ppb C
|
Operator Interface
|
Color LCD touchscreen display with Windows®
CE-based software
|
Operating Modes
|
Standalone (Windows® CE), PC-controlled, or LAN/
LIMS network connectivity
|
Basic Software
|
Single instrument operation with data transfer to PC
|
Optional ATOC Software
|
Network LAN/LIMS operation for data collection,
management and custom report generation
|
Autosampler
|
88 position rotary autosampler designed to fit
directly underneath Aurora 1030C analyzer
|
Sample Injection
|
Manual syringe, sipper tube, autosampler, or multistream
at-line sampling module
|
Certification
|
CE, EMC: EN61326 / Safety: IEC 61010-11 2001
|
Reagents Required
|
Hydrochloric acid, rinsewater
|
Sample Injection Volume
|
10µL – 0.8 mL
|
Method TC
|
680 ˚C Pt catalyst, 900 ˚C non-catalyst packing
|
Method TIC
|
Acidification and sparging
|
Method TOC
|
NPOC, combustion of TIC-free sample, TOC–TIC
|
Furnace Temperature
|
Adjustable, 680 – 950 ˚C in 1 ˚C increments
|
Repeatability
|
3.0%
|
Accuracy
|
±2% FS or 2% relative, whichever is greater
|
Sample pathway
|
Color coded Teflon® tubing
|
Sample handling
|
Syringe with isolation loop to prevent contamination
|
Gas Supply
|
Zero-grade air, or O2 (99.998%)
|
Power Supply
|
Variable voltage, 100-240VAC, 50/60 Hz, 950W
|
Dimensions – Aurora 1030
|
42.5 cm H x 49.5 cm W x 41.9 cm D
(16.75 in. H x 19.5 in. W x 16.5 in. D)
|
Dimensions – Aurora 1030 + 1088 Autosampler
|
26.75 in. H x 19.5 in. W x 23 in. D
|
Weight – Aurora 1030 + 1088 Autosampler
|
15.4 kg (34 lbs.), 34.5 kg (76 lbs.) 1030W + 1088
|