121L Pharmacy Refrigerator Model P 130
121L Pharmacy Refrigerator Model P 130
Description
Pharmacy Refrigerators for the safe storage of pharmaceuticals, medications or live virus vaccines, at cold temperatures between +4°C and +15°C with multiple controls and warning safety features developed as a result of 40 years of experience in refrigeration technology. The B Medical Systems 121L Pharmacy Refrigerator | Model P 130 is available in Australia exclusively from Banksia Scientific.
Integrated multi-function monitoring electronics
Integrated B Medical Systems multi-function electronics, at easy access level in the door; offers an alarm system, setpoint security, battery backup, full display, and many other control features; compatible with °B Connected solution for remote data monitoring and recording
Lifetime warranty corrosion-free cabinet
• One-piece material cabinet with a lifetime warranty for zero corrosion, highest quality polypropylene for better insulation and no sharp edges to provide easier maintenance and cleaning.
• Smart ergonomy – Easier and safer handling because of heavier components placed at the bottom.
High storage capacity
High storage per square meter in terms of net volume, with a very versatile inner volume utility due to the multiple user-friendly rails allowing modularity.
Highest insulation value (organic glass door)
4-layer PMMA organic glass offering highest insulation, combined with reduced weight reducing efforts to close door and constraints on hinges, offering better longevity.
Specifications
- Gross / Net volume (l) 121/106
- Set temperature (preset) +5°C
- Set temperature (setting range) can be adjusted in steps of 0.1°C +4°C to +15°C
- Temperature cold / warm alarm limit +3°C / +7°C
- Hold over time (+5°C to +10°C) 1 h 27
- Climate class (ambient temperature range) SN (+10°C to +32°C)
- Defrosting technique Natural
- Refrigerant type R600a
- External dimensions H x W x D (mm) 830 x 595 x 695
- Inner dimensions H x W x D (mm) 630 x 475 x 470
- Supply voltage (V) 220-240
- Frequency (Hz) 50/60
- Power (W) 200
- Energy consumption (kWh/24h) 0.86/0.92
- Heat emission (Kcal/h) 25
- Compressor running time (%) 26
- Noise level (dB(A)) (at 1m height & 1m distance) 35/36