Diamed Lab Supplies Inc

Diamed Lab Supplies Inc

5900 Ambler Dr #4, Mississauga, ON L4W 2N3, Canada

Amerex HV-Series Portable Top Loading Autoclaves

Amerex HV-Series Portable Top Loading Autoclaves

Amerex HV-Series Portable Top Loading Autoclaves

Features:

  • Space-saving design with lid opening upwards
  • Programmable auto-start for initiation of a cycle up to 1 week later
  • Dual-sensing lid interlock prevents lid opening when chamber is pressurized or when temperature exceeds 97° C
  • Automatic air purging by a microprocessor prevents over-pressurization during a cycle
  • One-touch lever to easily open and close chamber lid
  • Programmable rate of steam exhaust
  • Door-closure sensor provides a prompt and the cycle will not start unless the door is closed properly
  • In-process display shows a status and progress of sterilization cycle
  • Agar sterilization mode for rapidly melting agar and maintaining it at liquid temperature
  • Unique accessory: Floating (load) sensor to monitor load temperature.
  • Sterilization timer is activated only when load temperature reaches set temperature
  • Agar melting mode for rapidly, melting agar and maintaining it at liquid temperature, or for a hot and faster start-up
  • Memory backup system retains user-programmed parameters even in case of power failure
  • Not intended for medical use

Specifications:

 

AMEATHV-25

AMEATHV-50

AMEATHV-85

AMEATHV-110

Temperature

Operating Range

105º to 126º C

105º to 135º C

Display Range

128° C

137º C

Warming Range

45° to 60° C

Agar Dissolution

Heating at 60º to 100º C for up to 1 hour (temp. and time programmable), followed by 45º to 60º C (programmable for up to 20 hours to prevent coagulation)

Gauge Pressure

Operating Range

3 to 22 psi (0.02 to 148 MPa)

3 to 30 psi (0.02 to 206 MPa)

3 to 30 psi (0.21 to 0.02068 MPa)

Maximum Allowable

27 psi (0.186 MPa)

38 psi (0.26 MPa)

Timer

Sterilization

1 to 250 minutes

Delayed Auto-Start

Program Autoclave to start 1 minute to 7 days later

Dimensions

Overall (W x D x H)

19 x 18 x 37.4 inches,
48 x 46 x 95 cm

20 x 21 x 41 inches,
51 x 53 x 104 cm

26 x 25.6 x 39.4 inches, 66 x 65 x 100 cm

26 x 25.6 x 46.5 inches,
66 x 65 x 118 cm

Chamber (D x H)

25 litres,
9.4 x 21.6 inches,
24 x 55 cm

50 litres,
11.8 x 28 inches,
30 x 71 cm

85 liters,
16.5 x 24.2 inches, 42 x 61.5 cm

110 liters,
16.5 x 31.3 inches,
42 cm x 79.5 cm

Safety Devices

Dual-sensing lid interlock, over-pressure power cutoff, over-temperature power cutoff, low-water power cutoff, mechanical pressure relief valve, microprocessor-controlled pressure relief valve, lid-closure detector, over-current detector and circuit breaker

Programs

1. For Agar

Heating -> Sterilization -> Exhaust (from no exhaust to slow, programmable) -> Warming -> Completion

2. For Liquids

Heating -> Sterilization -> Exhaust (from no exhaust to slow, programmable) -> Completion

3. For Solids

Heating -> Sterilization -> Exhaust (fast, default) -> Completion

4. For Melting Agar or faster Start-Up

Rapid melting at 60° C to 100° C (temperature and time programmable) -> Maintain at 45° to 60° C (temperature programmable) -> Completion

Accessories

Stainless Steel

Mesh wire baskets, buckets (pails), and test tube rack

Factory or Field Installed

Floating (load) Sensor, Auto Water Supply Unit & Digital Printer

Floating (load) sensor, digital printer, and data logger

Power

Source

120 V, 50 / 60 Hz, (220 V / 240 V available on request)

208/220/240 V, 50/60 Hz, Single Phase

Consumption

1,500 Watts (1,500 VA)

2,000 Watts (2,000 VA)

3,000 Watts (3,000 VA)

4,000 Watts (4,000 VA)

Shipping Weight

160 lb, 73 kg

180 lb, 82 kg

282 lb, 128 kg

308 lb, 140 kg

Warranty

1 Year

Product Enquiry