P L Tandon & Co

P L Tandon & Co

205-B, Apra Plaza Commercial Complex, Road No.44, Pitampura, New Delhi-110034, (India)

Medical Waste Sterilizer (PLT-105 B)

Medical Waste Sterilizer (PLT-105 B)

Medical Waste Sterilizer (PLT-105 B)

APPLICATIONS OF MEDICAL WASTE STERILIZER

  • Robust and rigid construction.
  • Triple Walled Medical Waste Sterilizer (Rectangular) Specially designed for Sterilization of Regulated Bio-Medical Hospital Waste and related spores. It is based on the principle that microbiological action of saturated steam at elevated temperature is rapid and through.
  • Triple walled Medical Waste Sterilizer Rectangular is ideal tool for Hospitals & Nursing Homes.

SALIENT FEATURES & CONSTRUCTION OF MEDICAL WASTE STERILIZER

  • These Sterilizers work on the principle of downward displacement of air which is the most economical method of obtaining sterilization.
  • Triple walled with inner chamber and separate boiler is made up of Heavy gauge non magnetic S.S. Sheet (304/316 grade) argon-arc welded and jacket is made out of rolled steel.
  • Boiler Made of thick stainless steel fitted under the sterilizer. High grade flanged type heating elements are fitted from front side for easy replacement/repair. Fitted with water gauge glass for water level, safety valve, water inlet and drain valves provided with Low Water Level cut off device.
  • Whole unit horizontally mounted on a sturdy, heavy M.S. duly painted or Stainless steel tubular stand.
  • Lid (single piece door) made out of rolled steel lined with S.S. sheet on both sides tightened by Radial Locking System which cannot be opened when under pressure. The self-locking device automatically disengages the thread mechanism when the sterilizer is under pressure to preclude any eventuality of accident.
  • Sealing of Lid by Neoprene Rubber Gasket.
  • Fitted with shelf locking safety device.
  • Sterilization cycle is controlled by single point multi port valve fitted at front top. Pressure gauges are fitted to read jacket & chamber pressure on the top A moisture trap is fitted in the chamber discharge line to absorb the condensate automatically to prevent moistening in the chamber.
  • Chamber Insulated by Mineral/Glass Wool to minimize the heat losses and is covered by polished stainless steel sheet.
  • Fitted with pressure gauge, safety valve and steam release valve.
  • Fitted with plug screen to prevent the discharge line from choking.
  • Drain valve is fitted at the side bottom for easy draining/cleaning.
  • Sterilizing Temperature up to 1340C.
  • Sterilizing pressure : 1.2 to 1.5 kg/cm2 (15 psi to 22psi).
  • Hydraulically Tested up to 40 psi.
  • Fitted with Digital Temperature Indicator.
  • Fitted with Automatic Vacuum Breaker to break the vacuum in case of formation of vacuum due to steam condensation.
  • Fitted with Automatic Low Water Level Cut-off Device for Prevention of Heaters.
  • Supplied with Heavy Duty Vacuum Pump (Optional) for vacuum steam pulsating model.
  • Fitted with Automatic Pressure Control Switch to economizes power consumption by reducing frequent opening of the steam release valve. It cuts off the power supply to heaters when the set pressure is achieved and re-energizes the heating elements when pressure falls below the set point.
  • Control panel fitted with Pressure control switch, Main switch, Pilot lamps, relays etc. and mounted on the stand of the sterilizer for easy operation.
  • Voltage: 440 volts, 3 phase A. C. Supply

PARAMETERS

MWS - 1

MWS - 2

MWS - 3

Inner Dimensions Dia * Depth

400 x 600 mm

400 x 1100 mm

500 x 1200 mm

Inches Conversion

16" x 24"

16" x 44"

20" x 48"

Working Load

12.0 KW

12.0 KW

15.0 KW

Working Pressure Gauge

10-22 psi

Pressure Gauge

0-30 psi

Lid Quality

Heavy mild steel sheet lined with Stainless steel

Lid Tightening

Radial locking system worked by handle

Gasket

Made of Neoprene Rubber

Sterilizing pressure

1.2 kgf / cm (15psi)at 121°c

Pressure Gauge

0-2.1 kgf / cm² (30 psi)

Operating pressure

From 15 to 22psi (Adjustable)

Inner Pressure Tank

Stainless Steel (SS-304)

Basket

Made of Stainless Steel

Water Level Indicator

Fitted on the outer chamber

Boiler - Hydraulically Tested

Upto +40 psi

Temperature Display

By Digital Temperature Indicator

Vacuum

By Heavy Duty Vacuum Pump (Optional)

Accuracy

+ 3 psi

Safety Device

Spring Loaded Type

Water Draining System

Hand Operated Valve

Heating

BY ISI Marked Immersion Heaters

Steam & Vacuum Release

By Valve

Power Supply

440 volts, 3 phase A.C. Supply

OPTIONAL FEATURES OF MEDICAL WASTE STERILIZER

Temperature Recording

Thermograph with / without digital display - Circular Type

Micro Processor controller

With automatic Digital minute timer and alarm for cycle at the end.

PLC Controller

With 32 characters (2 x 16) LCD display for automatic purging of stale air sterilizing hold time for sterilizing period and cycle and with automatic exhaust pressure / recording & monitoring batch wise data.

S.S. Carriage

Available at extra cost

M.S. Trolley

Available at extra cost

Additional Tray/Shelves

Available at extra cost