The new carbon / sulphur analyser ELEMENTRAC CS-i was developed for the accurate and safe analysis of carbon and sulphur in inorganic samples. The analyser is equipped with a powerful induction furnace for sample combustion.

Up to four highly sensitive infrared (IR) cells allow the analyser to determine both high and low carbon and sulphur concentrations in only one measurement run. The measuring range of each cell may be adapted to the users specific requirements to ensure optimum measuring conditions for each application.

The analyser is supplied with the new comprehensive and convenient ELEMENTS software, which features statistics, groupings, reports, diagnosis tools and many additional functions for ELEMENTRAC analysers.

Advantages

  • simultaneous carbon and sulphur determination with minimal sample preparation
  • wide range of inorganic materials can be analysed
  • freely selectable configuration of each IR cell
  • optional gold IR path for analysis of halogen or acid containing samples
  • control of induction performance provides more precise analysis of samples with low melting temperature
  • powerful (2.2 kW) induction furnace for temperatures above 2,000 °C
  • NEW: automatic vacuum cleaning system ensures higher measurement precision and stability
  • heated and easily accessible dust trap allows for improved sulphur detection
  • optimized catalyst reactor permits more accurate carbon detection
  • NEW: ELEMENTS software with comprehensive analysis and diagnosis tools (supporting data and application export, comment fields, and many more)
  • single and multipoint calibration
  • low maintenance

robust design allows usage in production control and laboratory

Features

Measured elements

carbon, sulphur

Samples

inorganic

Furnace alignment

vertical

Sample carrier

ceramic crucibles

Field of application

construction materials, engineering / electronics, geology / mining, glass / ceramics, steel / metallurgy

Furnace

induction furnace, above 2,000 °C

Detection method

solid state infrared absorption

Number of IR cells

1 – 4

Material of IR path

aluminium (optional gold)

Typical analysis time

40 seconds (nominal)

Chemicals required

magnesium perchlorate, platinum-based catalyst, sodium hydroxide

Gas required

compressed air (4 bar/ 60 psi)
oxygen 99.5 % pure (2 – 4 bar / 30 – 60 psi)

Power requirements

230 V, 50/60 Hz, max. 15 A, 3450 W

Dimensions (W x H x D)

52 x 84 x 75 cm

Weight

~ 150 kg

Required equipment

balance (resolution 0.0001g), monitor, PC

Optional accessories

Autoloader (for 36 crucibles), carrier gas purification, halogen trap, HTF-540 pre-heating furnace, vacuum cleaner