ELEMENTRAC OH-P
The ELEMENTRAC OH-p determines oxygen and hydrogen in inorganic samples by inert gas fusion in an impulse furnace with temperatures in excess of 3,000 °C.
The ELEMENTRAC OH-p guarantees precise and fast sample analysis. The analyser covers a wide range of applications such as metal, ceramics and other inorganic materials.
The ELEMENTRAC OH-p can be supplied with up to two infrared cells with different path lengths, accommodating both high and low level oxygen analyses. Hydrogen concentrations are determined in the ELEMENTRAC OH-p by a robust and sensitive thermal conductivity cell.
Product Advantages
Features
Measured elements
hydrogen, oxygen
Samples
inorganic
Furnace alignment
vertical
Sample carrier
graphite crucibles
Field of application
ceramics, engineering / electronics, steel / metallurgy
Furnace
electrode impulse furnace (max. 8,5 KW*), temperatures in excess of 3,000 °C
Detection method
solid state infrared absorption for oxygen, thermal conductivity for hydrogen
Typical analysis time
120 – 180 s
Chemicals required
magnesium perchlorate, Schuetze reagents, sodium hydroxide
Gas required
compressed air, nitrogen 99.995 % pure, all gases with (2 – 4 bar / 30 – 60 psi)
Power requirements
3~ 400 V, 50/60 Hz, max. 8,500 W
Dimensions (W x H x D)
57 x 77 x 63 cm
Weight
~ 161 kg
Required equipment
balance (resolution 0.0001g), monitor, PC
Optional accessories
carrier gas purification, external chiller, gas calibration unit
–
* limited to 6.8 kw in application settings