Sysmex

Sysmex

Ecustraat 11 4879 NP Etten-Leur The Netherlands

R&D and Process Development Freeze Dryer

R&D and Process Development Freeze Dryer

R&D and Process Development Freeze Dryer

  • SMART Freeze-Dryer Technology helps automate and shorten the development of freeze-drying cycles
  • Clean room or isolator integrable
  • Configured with instrumentation for complete process information - isolation valve, capacitance manometers, Pirani gauge, 16 thermocouples, Allen Bradley PLC
  • Built as a ControLyo Controlled Nucleation Technologyready system which allows for adding ControLyo at any time
  • Optional LyoFlux TDLAS and Tempris

The LyoStar4 is designed to meet the needs of R&D professionals. Our freeze drying equipment is used for applications ranging from process development for scale-up to product processing for stability studies to formulations work. Models are available in stoppering and bulk configurations.

The LyoStar is an R&D and process development freeze-dryer that provides unmatched process accuracy, reliability and unique ease-of-use features. The LyoStar features RTD-controlled shelf temperature accuracy of +/- 0.5°C or better and vacuum control within 0.1% of set-point. Its robust 5.5 hp cascade refrigeration system enables shelf pull down from ambient to –40°C in less than 25 minutes, and the ultra-reliable scroll compressors feature just 3 moving parts.

Specifications

Lowest Shelf Temperature

≤ -70°C

Shelf Temperature Control Range

-70 to 60°C

Shelf Temperature Control Range Tolerance

± 0.5°C

Shelf Pull-Down from 25°C to -40°C

≤ 25 minutes

Lowest Condenser Temperature

≤ -85°C

Maximum Condenser Capacity

≥ 30 L

Condenser Surface Area

0.548 m2 (850 in2)

Condenser Pull-Down from 20°C to -45°C

≤ 10 minutes

Number of Compressors

2

Compressor Horsepower (high-stage / low-stage)

3.5 hp / 2 hp

System Refrigerant (high-stage / low-stage)

R449A / R508B

Vacuum Time to 100 Millitorr

≤ 25 minutes

Vacuum Rate of Rise (per hour)

≤ 40 μbar (30 mT)

Volume-Based Leak Rate

0.0019 mbar-L/sec(0.0014 Torr-L/sec)

Lowest System Vacuum

≤ 13 μbar (10 mT)

Vacuum Level Control Range

26 to 667 μbar (20 to

500 mT)

Vacuum Level Control

± 7 μbar (± 5.0 mT)

Temperature Uniformity§

± 1.0°C