BMB-III-Laminar-S-1,2 PROTECT VIS-À-VIS Class III
The cabinet is designed for operators working simultaneously opposite one another.
The system of monitoring with the audio-visible indication system warns operator about the failure of working modes in the cabinet.
Application of the Biological Safety Cabinet Class III:
Microbiological safety cabinet BMB-III-“Laminar-S”-1,2 PROTECT VIS-À-VIS Class III
Main Parameters and Dimensions
Article
1R-C.002-12
Maximum dimensions of the cabinet with pass box* /WхDхH/ (dimensions without stand are in parenthesis),mm
1720х880х1940(1270)
Dimensions of the work chamber /WхDхH/, мм
1040х655х685
Maximum net weight of the cabinet with stand,kg
350
Maximum input power without built-in electric sockets, W
450
Maximum input power,W
1450
Main Characteristics
Installation work chamber air cleanliness class for suspended particle (aerosol) concentration as per ISO 14644-1
- for particles of 0.5µm and more
ISO 5
- for particles of 5.0µm and more
ISO M (20; ≥5µm); LSAPC
Class of the cabinet as per EN 12469, NSF/ANSI 49
III
Minimum negative pressure in the work chamber,Pa
200
Constantly maintained (operating) negative pressure in the work chamber,Pa
250
Supply air filtration
Two-stage (G4, Н14)
Exhaust air filtration
Two-stage (Н14, H14)
Minimum illumination of the work zone (integral value determined along the whole area of the work zone),lx
2000
Recirculation, %
no recirculation
Maximum noise level at 1 m distance from the cabinet, dB
58*
* Уровень з** level of noise measured as per DIN EN ISO 11201 in free sound field over the sound-reflecting surface (noise level in real operating environment depends on the dimensions of the operating site as well as on the total background noise and may vary by 3-4 dB(A)). Uncertainty coefficient: k = 2 dB