BMB-III-Laminar-S-1,8 PROTECT CLASS III
The model design allows using those cabinets for work with deadly bacteria and viruses.
If connected to the individual active exhaust system, the cabinet could be used for protection from small amounts of toxic chemical substances and radionuclides.
The system of monitoring with the audio-visible indication system warns operator about the failure of working modes in the cabinet.
Key Benefits
The cabinet article 2Е-С.001-18 has 4 glove ports which allow two operators working simultaneously. The cabinet is produced upon request.
Microbiological safety cabinet BMB-III-"Laminar-S."-1,8 PROTECT Class III
Article
2E-C.001-18
Dimensions of the cabinet with the pass box* (WxDxH) / (dimensions without stand are in parenthesis), mm
2320х750х1940(1270)
Dimensions of the working chamber (WxDxH), mm
1740х640х675
Maximum net weight of the cabinet without stand,kg
375
Maximum net weight of the stand,kg
25
Power consumption (without the built-in outlets load), W, not more than
460
Maximum input power of the cabinet, W
1460
* dimensions do not account for outstanding supports
Main characteristics
Installation work chamber air cleanliness class for suspended particle (aerosol) concentration as per ISO 14644-1
-for particles of 0.5µm and more
ISO 5
-for particles of 5.0µm and more
ISO M (20; ≥5µm); LSAPC
Cabinet class according to ЕN 12469, NSF/ANSI 49
III
Class of the installed НЕРА-filters according to EN 1822-1
H14
Prefilter Class according to EN 779
G4
Minimal negative pressure in the working chamber, Pa, not less than
200
Permanently provided working negative pressure in the working chamber, Pa, not less than
250
Airflow rate through the supply HEPA filter, m3/h, not less than
400-450
Air inflow filtering
two-stage (G4, H14)
Air outflow filtering
two-stage (Н14, H14)
Illuminance level in the working zone, lux, not less than
2000
Air recirculation rate in the cabinet, %
no recirculation
Noise level at 1m distance from the cabinet, dB, not more than
59*
* level of noise at Main Operation mode measured as per DIN EN ISO 11201:2010 in free sound field over the sound-reflecting surface (noise level in real operating environment depends on the dimensions of the operating site as well as and on the total background noise and may vary by 3-4 dB(A)). Indeterminacy: k = 2 dB(A).