LAMSYSTEMS

LAMSYSTEMS

Magdeburger Str. 3, Wustermark bei Berlin, 14641 Germany

BMB-III-Laminar-S-1,8 PROTECT CLASS III

BMB-III-Laminar-S-1,8 PROTECT CLASS III

BMB-III-Laminar-S-1,8 PROTECT CLASS III

The model design allows using those cabinets for work with deadly bacteria and viruses.

If connected to the individual active exhaust system, the cabinet could be used for protection from small amounts of toxic chemical substances and radionuclides.

The system of monitoring with the audio-visible indication system warns operator about the failure of working modes in the cabinet.

Key Benefits

  • The cabinet has a sealed PASS-THROUGH SLUICE GATE made of stainless steel.
  • A PULL OUT SLUICE TABLE TOP makes loading objects into the working chamber easier.
  • FRONT SASH can be lifted for equipment loading. It is made of laminated safety glass and equipped with two oval glove ports.
  • REMOVABLE OUTLET UNIT in the working chamber with self-powered shutoff function.
  • The design of the cabinet allows integrating several cabinets in one line for continuous technological process.

The cabinet article 2Е-С.001-18 has 4 glove ports which allow two operators working simultaneously. The cabinet is produced upon request.

Microbiological safety cabinet BMB-III-"Laminar-S."-1,8 PROTECT Class III

Article

2E-C.001-18

Dimensions of the cabinet with the pass box* (WxDxH) / (dimensions without stand are in parenthesis), mm

2320х750х1940(1270)

Dimensions of the working chamber (WxDxH), mm

1740х640х675

Maximum net weight of the cabinet without stand,kg

375

Maximum net weight of the stand,kg

25

Power consumption (without the built-in outlets load), W, not more than

460

Maximum input power of the cabinet, W

1460

* dimensions do not account for outstanding supports

Main characteristics

Installation work chamber air cleanliness class for suspended particle (aerosol) concentration as per ISO 14644-1

-for particles of 0.5µm and more

ISO 5

-for particles of 5.0µm and more

ISO M (20; ≥5µm); LSAPC

Cabinet class according to ЕN 12469, NSF/ANSI 49

III

Class of the installed НЕРА-filters according to EN 1822-1

H14

Prefilter Class according to EN 779

G4

Minimal negative pressure in the working chamber, Pa, not less than

200

Permanently provided working negative pressure in the working chamber, Pa, not less than

250

Airflow rate through the supply HEPA filter, m3/h, not less than

400-450

Air inflow filtering

two-stage (G4, H14)

Air outflow filtering

two-stage (Н14, H14)

Illuminance level in the working zone, lux, not less than

2000

Air recirculation rate in the cabinet, %

no recirculation

Noise level at 1m distance from the cabinet, dB, not more than

59*

* level of noise at Main Operation mode measured as per DIN EN ISO 11201:2010 in free sound field over the sound-reflecting surface (noise level in real operating environment depends on the dimensions of the operating site as well as and on the total background noise and may vary by 3-4 dB(A)). Indeterminacy: k = 2 dB(A).